Walk-In Environmental Test Chamber
The walk-in chamber is well suited for testing the storage, transport and usage of electronic products, auto components and raw materials under high-temperature and low-temperature environments.
Technical Parameters of Walk-In ChamberModel | WTH40 | WTH60 | WTH80 | WTH120 | WTH160 | WTH180 | WTH230 | WTH290 | WTH310 | WTH390 | WTH470 | |
Nominal volume (m³) | 3.9/3.4 | 6.3/5.5 | 7.8/6.8 | 11.7/10.2 | 15.6/13.6 | 17.5/15.3 | 23.3/20.4 | 29.2/25.5 | 31.1/27.2 | 38.9/34.0 | 46.7/40.8 | |
Temperature range | -60°C -100°C (-70°C -120°C /-70°C -150°C optional) | |||||||||||
Humidity range | 20%RH-98%RH/(20-85°C) | |||||||||||
Temperature fluctuation | ±0.5°C | |||||||||||
Temperature deviation | ±2°C | |||||||||||
Humidity deviation | ±3.0%RH(when humidity ≧75%); ±5.0%RH(when humidity≦75%) | |||||||||||
Heat-up time | 25°C to +100°C(120°C, 150°C optional) | |||||||||||
≦25MIN | ≦28MIN | ≦28MIN | ≦30MIN | ≦31MIN | ≦32MIN | ≦34MIN | ≦36MIN | ≦38MIN | ≦41MIN | ≦45MIN | ||
Cool-down time | +25℃→ rated lowest temperature 30min -180min (depending on the model and configurations) | |||||||||||
Internal size (mm) | W | 1800 | 2700 | 3600 | ||||||||
H | I type: 2100 II type: 2400 | |||||||||||
D | 900 | 1450 | 1800 | 1800 | 1800 | 2700 | 3600 | 4500 | 3600 | 4500 | 5400 | |
Temperature and humidity adjustment method | BT(H)C balancing method | |||||||||||
Environment temperature | +5°C to +35°C | |||||||||||
Protection system | Galvanized steel sheets, temperature resistant insulation material, SUS304 stainless steel composite panels | |||||||||||
Regulator | Heater | Nichrome fin heater | ||||||||||
Humidifier | Stainless steel heater, steam humidification or external humidification. Ultrasonic humidifier is optional | |||||||||||
Fan | Centrifugal fan | |||||||||||
Refrigerator | Mechanical compressor | |||||||||||
Refrigerant | R404a or R404a /R23 | |||||||||||
Cooling method | When the lowest temperature ≥-25℃, mechanical water-cooling compression. When the lowest temperature ≤ -40℃, duplex water-cooling mechanical compression. | |||||||||||
Controller | Display | 7.0 inch, TFT color LCD display | ||||||||||
Control method | Touchscreen micro-computer controller, integral windup PID | |||||||||||
Auxiliary functions | Malfunction alarm and possible solution, power-off protection, temperature protection, calendar timing, automatic diagnosis | |||||||||||
Water supply | Humidification system | Multistage reverse osmosis water filtering system or water with an electrical resistivity ≥500Ωm. | ||||||||||
Circulating cooling water | Pressure: 0.32~0.45MPa, water temperature ≤30℃ | |||||||||||
Power supply | AC 380V(±10%)V three-phase four-wire + ground wire / AC220V / AC110V 50HZ /60HZ | |||||||||||
Power(KW) | 20KW to 55KW(depending on the model and configurations) |
Notes:
1. Height: I type: 2100mm, II type 2400mm.
2. Above parameters are measured when the ambient temperature is +25℃ and circulating water temperature is +25℃.
3. The fluctuation parameter is tested in accordance with GB/T5170.2-2008 standard.
4. The above-mentioned products are designed with a humidifier.
5. The WT(H): WT series products have no humidity function.
GB/T2423.1-2008 | Environmental Testing - Part 2: Test Methods - Tests A: Cold |
GB/T2423.2-2008 | Environmental Testing - Part 2: Test Methods - Tests B: Dry Heat |
GB/T2423.3-1993 | Basic Environmental Testing Procedures for Electric and Electronic Products - Test Ca: Damp Heat, Steady State |
GB/T2423.4-1993 | Basic Environmental Testing Procedures for Electric and Electronic Products - Test Db: Damp Heat, Cyclic GB10589-89 Specifications for Low Temperature Test Chambers |
GB10589-89 | Specifications for Low Temperature Test Chambers |
GB10592-89 | Specifications for Low/High Temperature Test Chambers |
GB/T10586-89 | Specifications for Damp Heat Testing Chambers |
GJB150.3 | Environmental Test Methods for Military Equipment - High Temperature Test |
GJB150.4 | Environmental Test Methods for Military Equipment - Low Temperature Test |
GJB150.9 | Environmental Test Methods for Military Equipment - Damp Heat Test |
GJB899-90 | Reliability Testing for Qualification and Production Acceptance |
GJB1032-90 | Environmental Stress Screening Process for Electronic Products |
The duct structure of Walk-In environmental test chamber has gained a patent for invention. The duct structure ensures that the temperature in each testing area of the test chamber is uniform, thus guaranteeing higher test accuracy.
Oil separator and oil returning and shunting device: The oil contained in the high-pressure steam discharged by refrigeration compressor is separated to ensure safe and efficient operation of the refrigeration system. Oil returning and shunting device (Simplewell’s patented technology) is mainly used to ensure the long-term operation of compressor or operation in high temperature. It prevents compressor failure or shortening of compressor service life due to lack of oil.
Optional Configurations
1. Rotary dehumidifier/ polymeric membrane, with a lowest temperature/ humidity content of 5℃/5 %RH.
2. Built-in tempered glass door, easy to move
3. Water purification system, with automatic replenishment function
4. Recorder, save and print testing data
5. Ultrasonic humidifier/ high-pressure spray humidifier, providing continuous high humidity under extreme temperature
6. Dry air blow system, anti-condensation
7. Remote monitoring function, cellphone and computer supportive
Cooling Equipment