2FW-32 Quiet Dental Air Compressor
Request a Quote
- Ambient temperature: 5℃~ 40℃
- Relative humidity: ≤80%
- Voltage: 198V-242V
- Frequency: 49Hz-51Hz
- The compressor should be installed on a flat ground space, with an angle of inclination after installation of no more than 2°.
Shaft power KW | Drive motor power KW | Nominal volume flow M³/min | Specific power M³/min | Rotating speed r/min | Capacity L | Noise Db(A) | Weight Kg | Dimensions mm |
0.84 | 0.85 | 0.064 | 9.3 | 1380 | 32 | 75 | 26 | 400×400×630 |
- When in use, the dental air compressor should be placed on flat surface. The working environment should be as clear as possible, with an excellent ventilation, no dust, and low humidity in order to ensure a good heat dissipation, and ensure the air source remains clean and dry.
- During usage, it is normal for a small amount of moisture to accumulate inside the air compressor tank, making regular drainage a requirement. Draining once a week is usually enough. Before turning off the water, make sure the power is off, and the drain valve is open, allowing the liquid to drain from the tank. When done, tighten the bottom valve on the air tank in order to continue using it.
- During usage, the intake muffler requires cleaning once a week, because dust in the air will enter the machine. Turn the power off before cleaning and removing the muffler. Open the end cover on the muffler and remove the interior sponge. Simply clean the sponge using clean water and allow it to dry. Reinstall it inside the muffler to continue using it. During this process, a replacement sponge can be used if needed.