Description
HJ series vertical steam sterilizer is equipped with automatic protection devices against over-temperature and over-pressure. When the temperature is too high, the sterilizer will automatically cut off the heating power. When the pressure is too high, it opens the safety valve for automatic pressure release protection.
Application
The HJ series vertical sterilizer is equipped with a built-in heater, sterilization timer, temperature and pressure automatic control regulator, safety valve, steam release valve, temperature and pressure indicator, alarm for finished sterilization, and heating power automatic cut off.
This sterilizer features a reliable sterilization effect. It is easy and safe to use, and its performance saves energy. It is durable, and features an exceptional quality at a low price. It is ideal for use in health care, scientific research and other industries for sterilizing surgical instruments, dressings, containers, culture mediums and more.
Features of HJ series vertical steam sterilizer
- Automatic sterilization process program control: it is easy to use and requires no supervision
- Steam temperature control program: the steam temperature is controlled within 105~134℃. Different objects can be sterilized at different required temperatures
- With a timing function, the automatic countdown begins when the sterilization temperature reaches the selected value. The sterilizer will automatically release pressure and steam, as well as cut off the heating power after the sterilization. It then alarms to indicate that the program operation is finished.
- The electric heating element is an immersion electrical heating tube, with a high thermal efficiency. It has an AC220V voltage and a 1KW power.
- This sterilizer is equipped with an open type safety valve. Under normal circumstances, it remains closed. When the pressure exceeds the maximum allowable range, it will open and release steam, and close when the pressure returns to normal.
- When the internal pressure exceeds 0.24MPa, it is impossible to open the self-locking door.
- Equipped with an overheat protector, the sterilizer will automatically cut off the power supply when water is lacking, not added, or cut off.
- Equipped with overload and electricity leakage protectors, the power will be automatically cut off when the circuit ether leaks or goes into over-voltage.
- The entire machine is made using a high quality stainless steel that is corrosion resistant, easy to maintain and features a long service life.
HJ35 parameters
Model | LS-35HJ |
Sterilizing chamber volume | 35L |
Maximum working pressure | 0.22MPa |
Maximum operating temperature | 134℃ |
Temperature uniformity of sterilization chamber | ≤±1℃ |
Adjustment of timing | 0~60min |
Adjustment of temperature/pressure | 105~134℃/ 0.04~0.22MPa |
Power/Voltage | 2500W/AC220V 50Hz |
Safety performance | Water cut off overheat protection, current overload automatic cut off, automatic overpressure release |
The door cannot be opened when the internal pressure exceeds 0.027Mpa, and it has the safety interlocking function. |
Overall size (L×W×H) | 450×450×1010 (mm) |
Transport volume | 570×550×1150 (mm) |
Weight | Gross weight 72Kg |
HJ50 Parameters
Model | LS-50HJ |
Sterilizing chamber volume | 50L |
Maximum working pressure | 0.22MPa |
Maximum operating temperature | 134℃ |
Temperature uniformity of sterilization chamber | ≤±1℃ |
Adjustment of timing | 0~60min |
Adjustment of temperature/pressure | 105~134℃/ 0.04~0.22MPa |
Power/Voltage | 3000W/AC220V 50Hz |
Safety performance | Water cut off overheat protection, current overload automatic cut off, automatic overpressure release |
The door cannot be opened when the internal pressure exceeds 0.027Mpa, and it has the safety interlocking function. |
Overall size (L×W×H) | 480×490×1030 (mm) |
Transport volume | 590×590×1230 (mm) |
Weight | Gross weight 86Kg |
HJ75 Parameters
Model | LS-75HJ |
Sterilizing chamber volume | 75L |
Maximum working pressure | 0.22MPa |
Maximum operating temperature | 134℃ |
Temperature uniformity of sterilization chamber | ≤±1℃ |
Adjustment of timing | 0~60min |
Adjustment of temperature/pressure | 105~134℃/ 0.04~0.22MPa |
Power/Voltage | 4500W/AC220V 50Hz |
Safety performance | Water cut off overheat protection, current overload automatic cut off, automatic overpressure release |
The door cannot be opened when the internal pressure exceeds 0.027Mpa, and it has the safety interlocking function. |
Overall size (L×W×H) | 560×560×1120 (mm) |
Transport volume | 650×630×1280 (mm) |
Weight | Gross weight 100Kg |
HJ100 Parameters
Model | LS-100HJ |
Sterilizing chamber volume | 100L |
Maximum working pressure | 0.23MPa |
Maximum operating temperature | 134℃ |
Temperature uniformity of sterilization chamber | ≤±1℃ |
Adjustment of timing | 0~60min |
Adjustment of temperature/pressure | 105~134℃/ 0.04~0.23MPa |
Power/Voltage | 4500W/AC220V 50Hz |
Safety performance | Water cut off overheat protection, current overload automatic cut off, automatic overpressure release |
The door cannot be opened when the internal pressure exceeds 0.027Mpa, and it has the safety interlocking function. |
Overall size (L×W×H) | 540×560×1250 (mm) |
Transport volume | 680×630×1370 (mm) |
Weight | Gross weight 110Kg |